PLEASE NOTE: All prices are exclusive of GST │ Please call / email our sales team to get the updated product price and to check the stock availability before placing an order. If you do not find the product that you require or the prices are not available ($0) please send us an email/ call us so we can assist. *Sold out products can be made available upon request.
Cart 0
Powervar One Rackmount Medical UPM ABCE600-22IECR High Voltage

Powervar

Powervar One Rackmount Medical UPM ABCE600-22IECR High Voltage

$0.00

Model Input-Output
Voltage (VAC)
Load
Power
(VA)
Size (mm.)
(H x W x D)
Shipping
Weight
(kg.)
Related
Docs
ABCE600-22IECR 230-230 600 132 (3U) X 483 X 496 25.4 VIEW FILES

 

POWERVAR Uninterruptible Power Manager: What a difference a letter makes.

If you think a UPS is the ultimate in power protection, we’ve got some news for you: Our Security One Medical UPM, or Uninterruptible Power Manager, performs functions your old UPS never dreamed of. And now, it’s available in a convenient 3U rack mount form factor that meets the requirements of UL60601 and IEC60601.

  Automatically adjusts for line voltage changes
  Low distortion sine wave inverter
  When equipped with its Communications Manager software suite, it can even track you down in the event of a power outage
  Integrated low impedance isolation transformer, noise filter and surge diverter assures power disturbances will be limited to <10 volts normal mode and <.5 volts (one half volt) neutral to ground
  North American models are listed to the special safety requirements of UL60601 and cUL 22.2 No. 60601 and are suitable for patient vicinity application
  International models are listed to the special safety requirements of IEC60601 and CE marked
 
  User hot-swappable battery
  USB and Serial Communication Interfaces
  Minimum runtime - 6 min full load / 20 min half load
  FCC Class A (Class B Optional)
  Five year warranty on electronics and two year warranty on battery